Advanced Research System Stereomicroscope (Galileo Parallel Optics) SZX16

Major Specifications
- 1. Newly designed optics delivers crisp, stereoscopic images
- 2. Varied types of powerful, intelligent illumination
- 3. The pursuit of fatigue-free ergonomic design
- 4. Sophisticated digital imaging system

SZX16 standard configuration
Advanced Research System Stereomicroscope (Galileo Parallel Optics) SZX16 : Features (1)
1. Newly designed optics delivers crisp real stereoscopic images
[1] Exceptionally wide zoom ratio 16.4 : 1
[2] Newly designed optics provide a high resolving power in 900 pair lines per mm (1.1 um) and maximum aberration correction.
[3] New Apochromat object lenses deliver advanced optical performance and operability at the same time.
450 pair lines / mm (2.2 um) and Working distance (Free space) 60 mm are fulfilled in the 1X objective lens.
[4] Unprecedented image clarity at most frequently used middle magnification thanks to the optical performance improvement by 30 % compared the traditional equivalent model.
[5] Newly designed SDF (Super Depth of Focus) objective lenses are adopted.
Newly designed SDF (Super Depth of Focus) objective lenses achieve crisp, advanced quality stereoscopic viewing with minimal defocusing, by effectively eliminating astigmatism.
 |
 |
Exceptionally wide zoom ratio 16.4 : 1 |
2x objective lens delivers 900 lines/mm resolving power. |
2. Varied range of powerful, intelligent illuminations
[1] This LED transmitted light illumination stand provides four image contrast options in an ultra-slim design.
[2] For intelligent industry solutions, the LED 4-part ring illumination system, coaxial illumination system and fiber related illumination are available.

Slim LED transmitted illumination base configuration
(available in autumn 06)

Co-axial illumination system
LED 4-part ring illumination system (SZX2-ILR66)
Advanced Research System Stereomicroscope (Galileo Parallel Optics) SZX16 : Specifications
| Item |
Specifications |
Zoom
microscope bodies |
SZX16 |
SZX10 |
Zoom ratio:
16.4:1 (0.7x-11.5x)
Magnification indication:
0.7/0.8/1/1.25/1.6/
2/2.5/3.2/4/5/6.3/8/10/11.5 |
Zoom ratio:
10:1 (0.63x-6.3x)
Magnification indication:
0.63/0.8/1/1.25/1.6/
2/2.5/3.2/4/5/6.3 |
Zoom variable magnification system with parallel optical axis
Zoom drive system: Horizontal handle Click-stop for various zoom positions incorporated |
| Built-in AS zoom body |
| Objective mounting: screw mount |
Objective
*With WHN10x-H/
WHSZ10-H |
For SZX-ZB16 |
For SZX-ZB10 |
| Objectives |
N.A. |
W.D.
(mm) |
Objectives |
N.A. |
W.D.
(mm) |
| SDFPLFL0.3X |
0.045 |
141 |
DFPL0.5X-4 |
0.05 |
171 |
| SDFPLAPO0.5XPF |
0.075 |
70.5 |
DFPL0.75X-4 |
0.075 |
116 |
| SDFPLAPO0.8X |
0.12 |
81 |
DFPLAPO1X-4 |
0.1 |
81 |
| SDFPLAPO1XPF |
0.15 |
60 |
SZX-ACH1X |
0.1 |
90 |
| SDFPLAPO1.6XPF |
0.24 |
30 |
DFPLAPO1.25X |
0.125 |
60 |
| SDFPLAPO2XPFC |
0.3 |
20 |
SZX-ACH1.25X-2 |
0.125 |
68 |
| |
|
|
DFPL1.5X-4 |
0.15 |
45.5 |
| |
|
|
DFPL2X-4 |
0.2 |
33.5 |
| Eyepiece |
WHN10X-H F.N. 22
WHSZ15X-H F.N. 16
WHSZ20X-H F.N. 12.5
WHSZ30X-H F.N. 7 |
WHSZ10X-H F.N. 22
WHSZ15X-H F.N. 16
WHSZ20X-H F.N. 12.5
WHSZ30X-H F.N. 7 |
| Observation heads |
SZX2-TTR/SZX2-TTRPT: Tilting trinocular head
Convergence angle, Tilting angle:5?-45?, ray switcher: 2 ways (TTR: binocular100%, binocular50%, camera50%, TTRPT: binocular100%, camera100%)
Interpupillary distance adjustment: 52-76mm |
SZX2-TR30/SZX2-TR30PT: 30 degree trinocular head
Convergence angle, Tilting angle:30?, ray switcher: 2 ways (TR30: binocular100%, binocular50%, camera50%, TR30PT: binocular100%, camera100%)
Interpupillary distance adjustment: 52-76mm |
| - |
SZX-BI30: 30? binocular head Tilting angle:30? Interpupillary distance adjustment: 51-76 |
| - |
SZX-BI45: 45? binocular head Tilting angle:45? Interpupillary distance adjustment: 52-76 |
| - |
SZX-TBI: tilting binocular head Tilting angle:5-45?
Interpupillary distance adjustment: 51-76 |
| Focusing assembly: |
SZX2-FO: Focusing unit / focus: rack and pinion with roller guide (with torque adjustment ring for focusing), optional counter balance,
coarse handle stroke: 80mm, coarse handle stroke per rotation: 21mm, Load capacity: 0-10.0kg |
SZX2-FOF: Fine focusing unit / focus: rack and pinion with roller guide (with torque adjustment ring for coarse focusing), coarse and
fine coaxial handle, built-in counter balance, coarse handle stroke: 80mm, coarse handle stroke per rotation: 36.8mm, fine handle
stroke: 80mm, fine handle stroke per rotation: 0.77mm, load capacity: 2.7-15.0kg |
SZX2-FOFH: : Heavy-duty fine focusing unit / focus: rack and pinion with roller guide (with torque adjustment ring for coarse focusing),
coarse and fine coaxial handle, built-in counter balance, coarse handle stroke: 80mm, coarse handle stroke per rotation: 36.8mm,
fine handle stroke: 80mm, fine handle stroke per rotation: 0.77mm, load capacity: 8.0-25.0kg |
SZX-FOA2: Motorized focus unit / focus: rack and pinion with roller guide (with torque adjustment ring for coarse focusing),
focusing stroke: 75mm, motorized focusing speed coarse:1.5mm/sec fine:0.3mm/sec load capacity: 2.7-15.0kg |
| Stands |
SZX-ST: Stand / Pillar height: 270mm, base dimension: 284(W)x335(D)x31(H)mm, Stage clips are mountable, with stage adapter fixing screw holes |
| SZX2-STL: Large stand / Pillar height: 400mm, base dimension: 400(W)x350(D)x28(H)mm, Stage clips are mountable, with stage adapter fixing screw holes |
| Environment |
In-house use / altitude: till 2000 m high, temperature: 5 through 40 degrees Centigrade, humidity: Max 80% (at or below 31 degree
Centigrade, the figure varies in other temperature.) |
Transmitted illuminators specifications
| Item |
Specifications |
| SZX2-ILLT |
SZX2-ILLB |
SZX2-ILLK |
SZX2-ILLD |
| Light source |
LED |
6V30W Halogen 6V30WHAL PHILIPS 5761 (average lamp service life: approx. 100 hours) |
| Light intensity adjustment |
Continuously variable system |
| Effective illuminated area |
Brightfield: ?63mm Darkfield / Oblique: ?35mm |
?40mm |
Brightfield: ?40mm Darkfield: ?35mm |
| Built-in filter |
- |
LBD, ND6, ND25 one for each |
- |
LBD (inter-lock brightfield) |
| Add-on filter |
- |
- |
?45mm frosted filter (45LBD), provided |
- |
| Illumination mode |
Brightfield illumination
Oblique illumination
Darkfield illumination |
Brightfield illumination
Oblique illumination |
Brightfield illumination
Oblique illumination |
Brightfield illumination
Darkfield illumination |
| Contrast selection |
- |
2-step selection of High and Low |
- |
- |
| Cooling fan |
- |
Built-in |
The height of stage
(from desk surface) |
41mm |
82mm |
| Pillar height |
270mm |
| Weight |
Approx. 3.7kg |
Approx. 5.0kg |
Approx. 4.6kg |
Approx. 5.4kg |
| Power source |
AC adaptor |
Built-in trance power unit |
Reflected light illuminators specifications
| Type |
LED ring illuminator SZX2-ILR66 |
Ring light guide LG-R66 |
Dual ring light guide LG-DFI/DI |
Coaxial illuminator SZX2-ILLC16/10 |
| Features |
4-part LED ring illumination
4-part independent
ON/OFF available |
Bright, uniformly lit images
without glaring reflections or
obscuring shadows |
Flexible illumination for any
angle and position |
Bright high contrast coaxial illumination.
Effective for observing structure,
such as imperfections on metal surfaces,
patterns on IC or LCD |
| Specifications |
Rotary, mirror function, ESD
capability, RoHS compatibility,
Class 1 |
Minimum WD: 30mm
Installation diameter: 66mm
Flexible part: 1000mm
Attachment adapter*: SZX-LGR66
*No adapter required for SZX16-LGR66
*Unable to attach to SDFPLAPO2XPFC/
SDFPLAPO16XPF |
LG-DFI: Flexible part 900mm
Inter-lock part 500mm
LG-DI: Inter-lock part 500mm |
Magnification factor: 1.5x
Light guide: LG-DF
Flexible part 1000mm
1/4 wave plate |
Light source
specifications |
10-segment light intensity
volume, fan-less ventilation,
AC100-240V |
Type: LG-PS2
Functions: Light intensity control and lamp ON/OFF control by external signal (DCO-5V), mechanical
adjustment function Power consumption: 150W (350VA)
Rated voltage: 100-120V/220-240V 50/60Hz
Dimensions: 120(H)x120(H)x235(D)mm Weight: approx. 1.5kg |
| Option |
SZX-LGR66,
adapter for SZX10 |
LG-R66PL: Polarizer/analyzer
set for LG-R66 |
HILL301: spot lens
LG-FAD: ?25 filter adapter |
|